Can the ASPTC-L and a sonic anemometer be used to measure high-frequency temperature fluctuations needed to find sensible heat flux using the eddy-covariance technique?

No. The ASPTC-L was not designed to measure high-frequency temperature fluctuations. Rather, it was designed to measure the mean air temperature or mean air temperature gradients. To find sensible heat flux using the eddy-covariance technique, consider using a sonic anemometer and an FW05 Fine Wire Thermocouple.