How many strain gages can be hooked up per excitation channel using a CR1000 and multiple 4WFBS120 modules?

The CR1000 excitation channel current limit is 25 mA at 2500 mV. Using 120 ohm gages with multiple 4WFBS120 modules, only two gages can be hooked up for each excitation channel with 2500 mV excitation. If three gages per excitation channel are used, the excitation needs to be reduced to approximately 1750 mV.