Trübungsmesser / OBS-3A
RETIRED ›
This product is no longer available. Some accessories, replacement parts, or services may still be available.
OBS-3A System zur Bestimmung von Trübung und Temperatur
Services Available

Überblick

Der OBS-3A kombiniert unseren OBS-Sensor mit Sensoren für Temperatur, Druck und Leitfähigkeit.  Der Mini-Datenlogger, Batterien und Elektronik befinden sich in einem Gehäuse, das in bis zu  300m Tiefe verwendet werden kann, abhängig vom Drucksensor.

(OBS® is a registered trademark of Campbell Scientific.)

Lesen Sie mehr

Funktionen und Vorteile

  • Runs 1,500 to 8,000 hours on three D-cell batteries
  • Monitors sediment concentrations up to 5,000 mg/l and turbidity up to 4,000 NTUs
  • Uses the field-proven OBS® technology (U.S. Patent No. 4,841,157) to measure turbidity
  • Logs instrument depth, wave height, wave period, temperature, and salinity
  • Records 200,000 lines of data in flash memory
  • Programs set points and sampling times
  • Offers an optional 5-point sedimentation calibration

Bilder

Technische Beschreibung

The heart of the OBS-3A monitoring system is an OBS sensor for measuring turbidity and suspended solids concentrations. This sensor detects near infrared (NIR) radiation scattered from suspended particles.

A fast-response, stainless steel-clad thermistor monitors temperature. Pressure is measured with a semiconductor piezoresistive strain gauge, and conductivity is measured with a four-electrode conduction-type cell. Working depths of the pressure sensor are selected as an option. (See the Ordering Info on the web page.)

The monitor uses HydroSci software running under Windows XP, 7, and 8.

Spezifikationen

Maximum Submersion Depth 300 m (984 ft)
Connector MCBH-8-FS, wet-pluggable
Operating Temperature Range 0° to 35°C
Storage Temperature Range -20° to +70°C
Drift over Time < 2% per year
Drift over Temperature 0.05% per °C
Maximum Sample Size 2048
Sampling Rate 1 to 25 Hz (when connected to a PC)
Maximum Data Rate
  • 25 Hz (connected to PC)
  • 5 Hz (used autonomously)
Data Capacity 8 MB
Maximum Number of Data Lines 200,000
Maximum Concentration Range
  • Concentration measurement range and accuracy depend on the sediment type.
  • 0.4 to 5,000 mg/L (for mud)
    Mud is defined as D50 = 20 μm.
  • 2 to 100,000 mg/L (for sand)
    Sand is defined as D50 = 250 μm.
PC Interfaces RS-232/115 kbps, RS-485/115 kbps
Battery Capacity 18 Ahr
Maximum Battery Life 8,000 h
Infrared Wavelength 850 nm
Pressure Measurement Range 0 to 10, 20, 50, 100, or 200 m
Turbidity Measurement Range 0.4 to 4,000 NTU
Conductivity Measurement Range 0 to 65 mS/cm (40 PSU, o/oo)
Concentration Accuracy
  • Concentration measurement range and accuracy depend on the sediment type.
  • 2% of reading (for mud)
    Mud is defined as D50 = 20 μm.
  • 3.5% of reading (for sand)
    Sand is defined as D50 = 250 μm.
Pressure Accuracy ±0.5% of f.s. (where f.s. = 50, 100, or 200 dBar)
Turbidity Accuracy < 2%
Temperature Accuracy ±0.5°C
Conductivity Accuracy 1%
Diameter 7.6 cm (3.0 in.)
Height 36.2 cm (14.3 in.)
Weight 1.5 kg (3.4 lb) without batteries

Kompatibel mit

Field Cables

Campbell Scientific offers a choice of field cables used to connect the OBS-3A to a PC for system configuration. The cables differ in their length. The HydroSci support software can be downloaded

Downloads

HydroSci v.1.2.1 (41.7 MB) 09-10-2017

HydroSci is used to configure and extract data from the CRS45X Recording Submersible Pressure Transducer and the OBS-3A Turbidity and Temperature Monitoring System.

Watch the Video Tutorial: Using HydroSci software with water-level recording sensors.

Liste der Änderungen

HydroSci Chinese Language v.1.2 (434 KB) 30-01-2017

Adds Chinese language support for version 1.2 of HydroSci.

Note: Language support for the above listed version only.