A547 Interface für den CS547A- Leitfähigkeitssensor
Zubehör zum Leitfähigkeitsensor
wetter wasser energie gas flux & turbulence bauwerke boden

Überblick

The A547 is an interface that allows our dataloggers to measure a CS547A water conductivity probe.

Lesen Sie mehr

Funktionen und Vorteile

  • Kompatibel mit AM16/32 Multiplexern zur Abfrage von zusätzlichen Sensoren

Bilder

Technische Beschreibung

The A547 interface consists of blocking capacitors and bridge completion resistors housed in a metal case. Screw terminals are provided for connecting the datalogger, and a water conductivity probe or multiplexer.

Spezifikationen

Operating Temperature -15° to +50°C
Dimensions
  • 6.4 x 4.6 x 2.3 cm (2.5 x 1.8 x 0.9 in.)
  • 8.9 x 4.6 x 2.3 cm (3.5 x 1.8 x 0.9 in.) with mounts
Weight 45 g (2 oz)

Kompatibel mit

Please note: The following shows notable compatibility information. It is not a comprehensive list of all compatible products.

Datenlogger

Product Compatible Note
CR1000
CR200X
CR216X
CR3000
CR5000
CR800
CR850
CR9000X

Sensors

Product Compatible Note
CS547A-L

Additional Compatibility Information

Datalogger Considerations

Measurement of the CS547(A) via the A547 requires two excitation channels, a differential input for the conductivity measurement, and a single-ended input for the temperature measurement.

Multiplexers

The A547 is compatible with AM16/32-series multiplexers. When a multiplexer is used, the sensors connect to the COM terminals of the multiplexer and the multiplexer connects to the sensor terminals of the A547.

Cabling

Datalogger-to-A547

A six-conductor cable, three feet in length, is provided with each A547 purchase for connecting it to the datalogger.

A547-to-Sensor(s)

In single-sensor applications, the sensor leads connect directly to the A547's terminal strips. In applications with multiple sensors, most customers multiplex the sensors through a multiplexer. In this case, a shielded 4-conductor cable is used to connect the sensor terminals on the A547 with the COM terminals on the multiplexer. 

Enclosures

The A547 requires a desiccated, non-condensing environment. The A547's case has mounting flanges for attachment to the 1"-on-center hole grid of Campbell Scientific enclosures. Grommets and screws are provided to attach the flanges to the backplate of our enclosures.