The SR50ATH is an identical unit to the SR50A with the addition of an external temperature sensor and transducer heater. The temperature sensor requires a six-plate radiation shield. The addition of a heating element around the transducer prevents ice and rime from coating the transducer with minimal power requirements. The SR50ATH outputs a temperature-corrected distance reading, eliminating the need for further post-processing, as well as a separate temperature reading.
The SR50ATH was designed to meet the stringent requirements of measuring snow depths and uses a multiple echo processing algorithm to help ensure measurement reliability. The SR50ATH is compatible with all current Campbell Scientific data loggers, as well as many other data acquisition systems.
Please note: The following shows notable compatibility information. It is not a comprehensive list of all compatible products.
| Product | Compatible | Note |
|---|---|---|
| CR1000 (retired) | SDI-12 and RS-232. For RS-485, requires an MD485 interface or similar RS-485 to RS-232 interface. | |
| CR1000X (retired) | SDI-12 and RS-232. For RS-485, requires an MD485 interface or similar RS-485 to RS-232 interface. | |
| CR300 (retired) | SDI-12 and RS-232. For RS-485, requires an MD485 interface or similar RS-485 to RS-232 interface. | |
| CR3000 | SDI-12 and RS-232. For RS-485, requires an MD485 interface or similar RS-485 to RS-232 interface. | |
| CR310 | SDI-12 and RS-232. For RS-485, requires an MD485 interface or similar RS-485 to RS-232 interface. | |
| CR350 | SDI-12 and RS-232. For RS-485, requires an MD485 interface or similar RS-485 to RS-232 interface. | |
| CR6 | SDI-12, RS-232, and RS-485 | |
| CR800 (retired) | SDI-12 and RS-232. For RS-485, requires an MD485 interface or similar RS-485 to RS-232 interface. | |
| CR850 (retired) | SDI-12 and RS-232. For RS-485, requires an MD485 interface or similar RS-485 to RS-232 interface. |
| Product | Compatible | Note |
|---|---|---|
| GRANITE 6 (retired) | SDI-12 and RS-232. For RS-485, requires an MD485 interface or similar RS-485 to RS-232 interface. |
One single-ended channel per probe is required; an excitation channel can be shared by several probes.
| Measurement Time | < 1.0 s |
| Output Options | SDI-12 version 1.3, RS-232, RS-485 (output options selected by configuring internal jumpers) |
| Baud Rates | 1200 to 38400 bps (RS-232, RS-485 modes) |
| Power Requirements | 9 to 18 Vdc (typically powered by data logger’s 12 Vdc power supply) |
| Quiescent Power Consumption in SDI-12 Mode | < 1.0 mA without heater |
| Quiescent Power Consumption in RS-232/RS-485 Mode |
|
| Active Power Consumption | 250 mA typical without heater |
| Measurement Range | 0.5 to 10 m (1.6 to 32.8 ft) |
| Beam Acceptance | ~30° |
| Resolution | 0.25 mm (0.01 in.) |
| Accuracy | ±1 cm (0.4 in.) or 0.4% of distance to target (whichever is greatest). |
| Operating Temperature Range | -45° to +50°C |
| Temperature Accuracy |
|
| Compliance | CE compliant |
| Diameter | 7.5 cm (3 in.) |
| Length | 10.1 cm (4.0 in.) |
| Weight | 375 g (13.2 oz) without cable |
| Cable Weight | 250 g (8.2 oz) for a 4.57-m (15-ft) cable |
Heater |
|
| Heater Resistance | 75 ohm |
| Nominal Operating Voltage | 12 V (ac or dc) Use a properly conditioned low-noise power source. A noisy power source will affect operation of the sensor. |
| Maximum Rated Wattage | 3 W |
| Maximum Rated Voltage | 15 V (ac or dc) |
| Maximum Operating Temperature | 25°C (Turn the heater power off at temperatures above 25°C. This prevents damage to the sensor and reduces power consumption.) |
Maximum Cable Length |
|
| -NOTE- | Cable lengths greater than 60 m require a heavier gage wire if the power supply drops below 11 Vdc. |
| SDI-12 | 60 m (200 ft) |
| RS-232 | 60 m (200 ft) Baud rates ≤ 9600 bps |
| RS-485 | 300 m (984 ft) |